Files:Technical information and downloadable documents 2
DiniArgeo SPSW series single point stainless steel load cells
A BL TECH specialist will confirm the price, lead time and technical fit.
Description
Stainless steel load cells for off center loads, suitable for example to build CE-M approved scales with a single load cell with maximum dimensions of 500x400 mm.
C3 PRECISION CLASS
OFF CENTER LOAD COMPENSATED
MADE IN 17-4 PH STAINLESS STEEL
IP67 PROTECTION
MAX. CAPACITIES FROM 7.5 TO 200 kg
FOR PLATFORMS WITH DIMENSIONS UP TO 500 X 400 mm
SHIELDED CONNECTION CABLE
TEST CERTIFICATE AVAILABLE UPON REQUEST
● Minimum load cell verification interval: Vmin = EMax / 10.000-15.000
● Maximum number of verification intervals: nLC = 3000
● Full Scale Output: 2mV/V +/- 10%
● Temperature effect on zero:
from ±0,0093% F.S./10K to ±0,0140% F.S./10K
● Temperature effect on full scale output:
±0,0117% F.S./10K (-10°C/+20°C)
±0,0175% F.S./10K (+20°C/+40°C)
● Compensated Temperature Range: -10°C/+40°C
● Operating Temperature Range: -10°C/+50°C
● Creep at nominal load in 30 minutes: ± 0,01% F.S.
● Maximum tolerated excitation voltage: 15 VDC
● Input Resistance: 380 ± 15 Ohm
● Output Resistance: 359 ± 5 Ohm
● Insulation Resistance at 100V: >1000 MOhm
● Safe Overload: 150% F.S.
● Ultimate Overload: 300% F.S.
● Nominal displacement: < 0.5 mm
● 6-wire shielded cable, 3m long, 5mm diameter
The CCATEX certificate makes the load cell suitable for use in hazardous areas with protection methods according to:
● ATEX II 1G Ex ia IIC T6 (Ta -20÷+40°C) TX (Ta -20÷+65°C) Ga
● ATEX II 1D Ex ta IIIC TX°C (Ta -20÷+40°C) TX°C (Ta -20÷+65°C) Da IP65.
The certificate includes the manual and the ATEX CE conformity declaration (available in EN, DE, FR, and IT).
IMPORTANT: the certificate refers to the single load cell. For systems with more load cells, it is necessary to order one certificate for each load cell.